RespiratorsChapter 296-842, WAC |
Effective Date: 12/01/09 |
Required Procedures for Respiratory Protection Program |
|
| Use this medical questionnaire for medical evaluations |
|
| Follow these fit-testing procedures for tight-fitting respirators |
WAC 296-842-22010 |
| Follow procedures established for cleaning and disinfecting respirators |
WAC 296-842-22015 |
| Follow procedures established for seal checking respirators |
WAC 296-842-22020 |
Use this medical questionnaire for medical evaluations
• Use the medical questionnaire in
Table
10 when conducting medical evaluations.
|
|
Note: • You may use a physical exam instead
of this questionnaire if the exam covers the same information
as the questionnaire. |
TABLE
10 - Medical Questionnaire (English)
TABLE
10 - Medical Questionnaire (Spanish)
Follow these fit-testing procedures for tight-fitting respirators
Important:
• This section contains procedural requirements
that apply during actual fit testing.
• See WAC 296-842-15005 of this
chapter for fit-testing requirements that apply to your overall
program.
|
This section does not apply to employees who: • Voluntarily use respirators; |
(1) Follow the procedure in
Table
11 to choose a respirator for fit testing:
- (a) Prior to conducting fit tests;
and - (b) Any time your employee must select a different respirator such as when a previously selected respirator fails a test.
(2) Select and follow at least one of the following fit test procedures:
(a) Qualitative fit-test procedures:
- Isoamyl acetate vapor (IAA, banana oil) in
Table 12;
- Saccharine aerosol in
Table 13;
- Bitrex™ aerosol in
Table 14;
- Irritant smoke in
Table 15;
(b) Quantitative fit-test procedures:
- Ambient aerosol condensation nuclei counter such as the Portacount™, in
Table 16;
- Controlled negative pressure (CNP) such as the FitTester 3000™, in
Table 17;
- Generated aerosol in
Table 18.
(3) Make sure employees perform the appropriate
fit test exercises listed in
Table
19.
(4) Clean and maintain equipment according to the manufacturer’s instructions.
(5) Make sure during fit testing employees wear any safety equipment that could:
(a) Interfere with respirator fit;
and(b) Be worn in the workplace. For example, chemical splash goggles.
(6) Check, prior to fit testing, for conditions that may interfere with the respirator seal or valve functions. If you find such conditions, do not conduct fit testing for that individual.
|
|
Note: Examples of conditions that may interfere with the respirator seal or valve functions include:
|
(7) Follow the appropriate fit test exercises in Table 19 as indicated.
Follow procedures established for cleaning and disinfecting respirators
• Follow the procedures in Table 20 for cleaning and disinfecting respirators.
Follow procedures established for seal checking respirators
Make sure employees perform a user seal check as outlined in Table 21 each time the respirator is worn, to make sure the seal is adequate.
Important:
• User seal checks are not
a substitute for fit tests. See
WAC 296-842-22010
for fit test procedures.
• You may use a seal check procedure recommended by the
respirator manufacturer instead of the procedure outlined in
Table
21 if you can demonstrate the procedure is
based on a scientific study that, for example, demonstrates the
procedure effectively identifies respirators that fit poorly when
put on or adjusted.
